Development of a calorimetric sensor for medical application |
| |
Authors: | Ch. Jesús F. Socorro M. Rodríguez de Rivera |
| |
Affiliation: | 1. Departamento de Física, Universidad de Las Palmas de Gran Canaria, 35017, Las Palmas de Gran Canaria, Spain
|
| |
Abstract: | A calorimetric sensor has been developed for local and direct measurement of calorific dissipations in different parts of the human body. The constructed prototype has a detection surface of 36 cm2. The calibration of the sensor is based on a semi-empirical model that permits to simulate the operation of the device, making easier an operational functioning method. The device is modeled as a system with two inputs and two outputs. The inputs are the calorific power (W) that is intended to be measured and the power (W pid) that dissipates a resistance, keeping constant the thermostat temperature through the use of a PID controller. The outputs are the thermostat temperature (T pid) and the calorimetric signal (y) that provides the thermopile that is in contact with the body. |
| |