首页 | 本学科首页   官方微博 | 高级检索  
     检索      


New method for determining the concentration of ionic impurities in solvent polymeric membranes
Authors:Mathias Nägele  Ernö Pretsch
Institution:(1) Department of Organic Chemistry, Swiss Federal Institute of Technology (ETH), Universitätstrasse 16, CH-8092 Zürich, Switzerland
Abstract:A new method is proposed for ascertaining the amount of ionic impurities in solvent polymeric membranes of ion-selective electrodes. The method is based on determining the selectivity coefficient for ions of different valences as a function of the concentration of a lipophilized tetraphenylborate salt added to the membrane phase. Thus, the concentration of anionic impurities in commercially available poly(vinyl chloride) and of cationic ones in Tecoflex® (a polyurethane) was obtained as 0.063 ± 0.016 and 0.044 ± 0.006 mmol/kg, respectively.
Keywords:ionic sites  PVC  Tecoflex®    ion-selective electrodes
本文献已被 SpringerLink 等数据库收录!