首页 | 本学科首页   官方微博 | 高级检索  
     


Monitoring of fluid motion in a micromixer by dynamic NMR microscopy
Authors:Ahola Susanna  Casanova Federico  Perlo Juan  Münnemann Kerstin  Blümich Bernhard  Stapf Siegfried
Affiliation:NMR Research Group, Department of Physical Sciences, 90014 University of Oulu, P.O. Box 3000, Finland. suahola@sun3.oulu.fi
Abstract:The velocity distribution of liquid flowing in a commercial micromixer has been determined directly by using pulsed-field gradient NMR. Velocity maps with a spatial resolution of 29 microm x 43 microm were obtained by combining standard imaging gradient units with a homebuilt rectangular surface coil matching the mixer geometry. The technique provides access to mixers and reactors of arbitrary shape regardless of optical transparency. Local heterogeneities in the signal intensity and the velocity pattern were found and serve to investigate the quality and functionality of a micromixer, revealing clogging and inhomogeneous flow distributions.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏