首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of doxycycline in pharmaceuticals and human urine by micellar electrokinetic capillary chromatography
Authors:Rade Injac  Javor Kac  Samo Kreft  Borut Strukelj
Affiliation:(1) Institute of Pharmaceutical Biology, Faculty of Pharmacy, University of Ljubljana, Askerceva 7, 1000 Ljubljana, Slovenia
Abstract:Micellar electrokinetic capillary chromatography (MEKC) was performed at 25 °C and 30 kV (under a pressure of 15 mbar), using 30 mM borate buffer containing 60 mM sodium dodecysulfate (SDS) and 5% (v/v) methanol as background electrolyte (pH 9.0) to determine doxycycline. UV detection was at 350 nm. The method was shown to be specific, accurate (recovery was 100.3 ± 1.0%), linear over the tested range (correlation coefficient 0.9995) and precise (RSD <1.9%). The method was used to determine doxycycline in tablets, capsules and human urine after oral application.
Keywords:Doxycycline  Micellar electrokinetic capillary chromatography (MEKC)  Pharmaceuticals  Human urine