On alpha particle spectroscopy based on the over-etched track length in PADC (CR-39 detector) |
| |
Authors: | M. El Ghazaly |
| |
Affiliation: | 1. Department of Physics, Faculty of Science , Zagazig University , PO 44519, Zagazig , Egypt ghazaly2000@yahoo.com |
| |
Abstract: | In the current work, alpha particle spectroscopy is investigated experimentally by utilizing the over-etched track lengths in the CR-39 detector. CR-39 samples were exposed perpendicularly to alpha particles emitted from 241Am with an energy ranging from 0.5 to 5.5 MeV. CR-39 samples were etched in 6.25 N NaOH at (70±0.5)°C for different durations. The track diameter and track length were measured under an optical microscope. The results show that, the energy-over-etched track length calibration curve is monotonic, in other words, the over-etched track length is a monotonic function in alpha particle energy. On the other hand, the energy-diameter calibration curve is degenerated, i.e. alpha track diameter is non-monotonic function in alpha particle energy. These results suggest that the CR-39 detector could be used as a wide range alpha particles spectrometer using an energy-over-etched track length calibration curve. |
| |
Keywords: | poly allyl diglycol carbonate (CR-39) track-etching rate over-etched track length alpha particles spectrometry monotonic and non-monotonic function |
|
|