首页 | 本学科首页   官方微博 | 高级检索  
     


On alpha particle spectroscopy based on the over-etched track length in PADC (CR-39 detector)
Authors:M. El Ghazaly
Affiliation:1. Department of Physics, Faculty of Science , Zagazig University , PO 44519, Zagazig , Egypt ghazaly2000@yahoo.com
Abstract:In the current work, alpha particle spectroscopy is investigated experimentally by utilizing the over-etched track lengths in the CR-39 detector. CR-39 samples were exposed perpendicularly to alpha particles emitted from 241Am with an energy ranging from 0.5 to 5.5 MeV. CR-39 samples were etched in 6.25 N NaOH at (70±0.5)°C for different durations. The track diameter and track length were measured under an optical microscope. The results show that, the energy-over-etched track length calibration curve is monotonic, in other words, the over-etched track length is a monotonic function in alpha particle energy. On the other hand, the energy-diameter calibration curve is degenerated, i.e. alpha track diameter is non-monotonic function in alpha particle energy. These results suggest that the CR-39 detector could be used as a wide range alpha particles spectrometer using an energy-over-etched track length calibration curve.
Keywords:poly allyl diglycol carbonate (CR-39)  track-etching rate  over-etched track length  alpha particles spectrometry  monotonic and non-monotonic function