首页 | 本学科首页   官方微博 | 高级检索  
     


A specific enzyme electrode for urea
Authors:G G Guilbault  M Tarp
Affiliation:Chemistry Department A, Technical University of Denmark, 2800 LyngbyDenmark
Abstract:A truly specific, simple enzyme electrode is described for the assay of urea in blood serum. The sensor used is the newly developed air-gap electrode of R??i?ka and Hansen, and has advantages of speed of response and specificity over earlier enzyme electrodes for urea. Potassium, sodium and ammonium ions and other organic and inorganic species present in blood do not interfere. Linear curves are obtained from 2 · 10-2M to 1 · 10-4M urea with slopes close to Nernstian (about 0.90 pH/decade). Urea in blood was assayed with an accuracy of 2.2% and a precision of 2.0% with immobilized urease; only 3–5 min is required per assay. The electrode was used for a month and almost 500 assays with excellent results. Since the sensor never touches the sample solution, problems caused by blood components which block membrane pores are avoided.
Keywords:
本文献已被 ScienceDirect 等数据库收录!