Trace level vanadium determination using thermal lens spectrophotometry |
| |
Authors: | Manuel Caetano Leonardo Padrino T Hector Gutiérrez Alberto J. Fernández Jimmy Castillo |
| |
Affiliation: | (1) Escuela de Química, Facultad de Ciencias, Universidad Central de Venezuela, P.O. Box 47102, 1020a Caracas, Venezuela |
| |
Abstract: | Trace level vanadium determination is reported using a dual beam thermal lens spectrometer. The thermal lens was generated using an argon ion beam laser (pump beam) which was focused into a sample cuvette. The thermal lens signal (TLs) was monitored with a He-Ne laser beam and a photodiode detector. Multichannel averager software was developed for processing the transient TLs. The optimal set up, ensuring maximum sensitivity and linear calibration graphs was obtained using experimental design techniques. Under optimized conditions, the detection limits for aqueous and ethanol-water (2+3 v/v) and (4+1 v/v) vanadium complex solutions were, respectively, 0.0071 mg/l, 0.0065 and 0.0039 mg/l. |
| |
Keywords: | thermal lens vanadium analysis trace analysis spectrophotometry optimization |