Three dimensional magnetic resonance imaging by magnetic resonance force microscopy with a sharp magnetic needle
Authors:Tsuji S  Yoshinari Y  Park H S  Shindo D
Institution:Advanced Technology Division, JEOL Ltd., 3-1-2 Musashino, Akishima, Tokyo 196-8558, Japan. stsuji@jeol.co.jp
Abstract:An electropolished magnetic needle made of Nd(2)Fe(14)B permanent magnet was used for obtaining better spatial resolution than that achieved in our previous work. We observed the magnetic field gradient |G(Z)|=80.0G/microm and the field strength B=1250G at Z approximately 8.8 microm from the top of the needle. The use of this needle for three dimensional magnetic resonance force microscopy at room temperature allowed us to achieve the voxel resolution to be 0.6 microm x 0.6 microm x 0.7 microm in the reconstructed image of DPPH phantom. The acquisition time spent for the whole data collection over 64 x 64 x 16 points, including an iterative signal average by six times per point, was about 10 days.
Keywords:MRFM  ESR  ESR imaging  Permanent magnet  Magnetic needle