Monte Carlo generated spectra for QA/QC of automated NAA routine
Authors:K. R. Jackman  S. R. Biegalski
Affiliation:(1) Department of Mechanical Engineering, University of Texas at Austin, Nuclear Engineering Teaching Laboratory, R-9000, Austin, Texas 78712, USA
Abstract:A quality check for an automated system of analyzing large sets of neutron activated samples has been developed. Activated samples are counted with an HPGe detector, in conjunction with an automated sample changer and spectral analysis tools, controlled by the Canberra GENIE 2K and REXX software. After each sample is acquired and analyzed, a Microsoft Visual Basic program imports the results into a template Microsoft Excel file where the final concentrations, uncertainties, and detection limits are determined. Standard reference materials are included in each set of 40 samples as a standard quality assurance/quality control (QA/QC) test. A select group of sample spectra are also visually reviewed to check the peak fitting routines. A reference spectrum was generated in MCNP 4c2 using an F8, pulse-height, tally with a detector model of the actual detector used in counting. The detector model matches the detector resolution, energy calibration, and counting geometry. The generated spectrum also contained a radioisotope matrix that was similar to what was expected in the samples. This spectrum can then be put through the automated system and analyzed along with the other samples. The automated results are then compared to expected results for QA/QC assurance.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏