Monte-Carlo simulations of Al2O3 dosimetry in uniform MV photon beams: Influence of field and detector size |
| |
Authors: | Xuetao Wang Jinhan Zhu Shaowen Chen Qiang Tang Xiaowei Liu |
| |
Institution: | 1. School of Physics Science and Engineering, Sun Yat-Sen University, Guangzhou 510275, China;2. School of Electronic Engineering, Dongguan University of Technology, Dongguan 523808, China |
| |
Abstract: | This work was to evaluate the absorbed dose to different sizes Al2O3 dosimeter in uniform MV photon beams using Monte Carlo simulation. The absorbed dose ratio factor fmd of Al2O3 dosimeters was calculated. The incident beams included 4 MV–24 MV X-rays using Mohan's spectra with different field sizes. Results show that the maximum variation of fmd due to different dosimeter sizes is 5%. As the field diameter increases from 1 cm to 3 cm, the difference of fmd can be up to 8%. The results of this work show that the effect of dosimeter size and the field size can't be neglected when using Al2O3 dosimeter. |
| |