首页 | 本学科首页   官方微博 | 高级检索  
     检索      


The real-time gamma radiation dosimetry with TeO2 thin films
Authors:TK Maity  SL Sharma  G Chourasiya
Institution:1. Department of Physics and Meteorology, Indian Institute of Technology, Kharagpur 721302, India;2. MP & TS – RPAD, BARC, Anushakti Nagar, Mumbai 400094, India
Abstract:A detailed study of the effect of gamma radiation on the current–voltage characteristics of the TeO2 thin films of different thicknesses, prepared by thermal evaporation in a vacuum, has been carried out for a much wider range of the gamma radiation doses than made here-to-fore. Subsequently, for the thin films of different thicknesses at different applied voltages, the variations of the current density with dose have been obtained. The current density increases near linearly with the gamma radiation dose up to a critical radiation dose, a dose value higher for the thicker films and decreases thereafter. The sensitivities of these TeO2 thin films at different applied voltages have been found to be in the range 1.2–37.0 nA/cm2/μGy. Correspondingly, the detection limits have also been estimated and have been found to be in the range 0.22–2.16 mGy. Clearly, the TeO2 thin films have high potential for their use as real-time gamma radiation dosimeters in monitoring the gamma radiation doses under a variety of practical situations involving low level to high level of the doses.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏