An experimental study of the relative response of plastic scintillators to photons and beta particles |
| |
Authors: | A. Kumar A.J. Waker |
| |
Affiliation: | Faculty of Energy Systems and Nuclear Science, University of Ontario Institute of Technology, 2000 Simcoe Street North, Oshawa, Ontario, Canada L1H 7K4 |
| |
Abstract: | A scintillation counting system has been constructed with the use of BC-400 and EJ-212 series plastic scintillators along with a subminiature photomultiplier tube to investigate the effect of increasing plastic scintillator thickness on system-integrated counts. Measurements have been carried out using four different gamma sources with different energies ranging from 6 keV to 1.332 MeV and a Ni-63 beta source with a maximum energy of 66 keV. Scintillator thicknesses ranged from 10 μm to 2500 μm. The response of the system was determined by measuring the integrated counts as a function of scintillator thickness. These experimental findings were used to empirically determine the optimum thickness of scintillator material with which to build a low energy beta detector which discriminates against high energy gamma photons in a mixed radiation field environment. |
| |