首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Intensification of the Emission Signals Using the Reflection Mirrors Mounted to the Sample Holder of a Fluorimeter
Authors:Takao Itoh
Institution:(1) Graduate School of Integated Arts and Sciences, Hiroshima University, 1-7-1 Kagamiyama, Higasi-Hiroshima City 739-8521, Japan
Abstract:Effects of the reflection mirrors mounted to the sample holder of a fluorimeter have been investigated. Analytical expressions are given for the emission intensity measured as a function of the concentration or optical density of the sample, in case one or two reflection mirrors are attached beside the sample holder to intensify the emission signals. The emission intensity calculated as a function of the sample concentration agreed well with the experimental data. By mounting two reflection mirrors, the emission signal was intensified by a factor of near 3.5. However, the degree of intensification depended strongly on the sample concentration and the linearity between the sample concentration and the emission intensity deviated significantly with increasing the sample concentration.
Keywords:Signal intensification  Emission  Fluorimetry  Optical density  Reflection mirrors
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号