Polymer crystallization dynamics,as reflected by differential scanning calorimetry. Part 1: On the calibration of the apparatus
Authors:H. Janeschitz-Kriegl  H. Wippel  Ch. Paulik  G. Eder
Affiliation:(1) Abteilung Verfahrenstechnik, Institute of Chemistry, Linz University, Altenbergstraße 69, A-4040 Linz, Austria
Abstract:In this paper a method for the calibration of the heat transfer coefficient between pan and furnace is given. This (second) calibration is necessary in addition to the usual calibration of the temperature scale. Indeed, with increasing cooling rates as required for kinetic measurements, the finite heat flow resistance between pan and furnace becomes evident anyway. We also propose to enlarge this resistance deliberately, in order to separate the time scales of the control system and of the exponential return of the heat flow curve to the base line, as occurring after completion of the phase transition. Only in this way can the heat transfer coefficient be determined with some accuracy. Another advantage of a lowered heat transfer coefficient will be treated in a third paper. It enables an approximate treatment of polymer crystallization kinetics.
Keywords:Differential scanning calorimetry  calibration  sample size  pan quality  internal heat flow resistance  apparent heat transfer coefficient