Department of Chemistry, Texas A and M University, College Station, TX 77843-3255, USA
Abstract:
The most effective way to obtain high quality vapor-phase Raman spectra is to heat the samples to increase their vapor pressure. Many samples can be heated to 350 °C and higher without decomposition. We have designed a simple Raman cell to allow these high temperature studies to be carried out. The high-temperature Raman spectra of nine molecules will be presented and discussed. Most of these are non-rigid molecules containing aromatic rings for which vibrational potential energy surfaces have been determined from their spectra. Two molecules (p-cresol and 3-methylindole) are model compounds for amino acids and their vapor-phase spectra are characteristic of environments with no hydrogen bonding.