Abstract:In the present work, a comparative study is made among nuclear methods for235U/238U ratios determination: activation analysis followed by high-resolution gamma-ray spectrometry, delayed neutron counting, passive gamma-ray and alpha spectrometry. Activation analysis followed by high-resolution gamma-ray spectrometry yielded a relative standard deviation down to 0.1% and a relative error of about 1% for standards of uranium enriched to 14%. Passive gamma-ray spectrometry using Ge(Li) detectors yielded a relative error down to 0.1% for enriched uranium and values even lower for the standard deviation. Passive gamma-ray spectrometry using Low Energy Photon Detector (LEPD) yielded a precision of 0.2% and a still better accuracy for enriched standards. In the case of alpha spectrometry, a relative error down to 0.5% and a precision of about 1% were obtained, also for enriched uranium standards. Delayed neutron counting allowed a relative standard deviation of about 7% and a relative error of about 2%, for standards of depleted uranium.