Development and characterization of an electrospray ionization ion trap/linear time-of-flight mass spectrometer
Authors:Randy W. Purves  Liang Li
Affiliation:1. Department of Chemistry, University of Alberta, T6G 2G2, Edmonton, Alberta, Canada
Abstract:On-line analysis of compounds from solution has been greatly facilitated by the advent of electrospray ionization mass spectrometry (ESI-MS). Although quadrupole mass analyzers are most commonly used with ESI at present, time-of-flight (TOF) mass spectrometers offer several potential advantages including high data acquisition rates, which are desirable for fast separation techniques. One method of coupling ESI and TOF uses an ion trap for temporary storage and accumulation of the electrosprayed ions prior to TOF mass analysis. Previous studies have not fully addressed the effects of several key variables on the analytical capabilities of this type of instrument. In this study, the characterization of an ion trap/linear TOF instrument for ESI is described. The behavior of various analytes is divided into two separate groups; each one is found to have its own optimal set of operating conditions. The reasons for the observed differences between groups are explored. Issues relevant to mass resolution, sensitivity, mass range, mass-to-charge ratio discrimination, and mass measurement accuracy are addressed. Finally, it is suggested that the analytical capability of this type of instrument could be significantly improved by changing the ion optics from the existing focusing lenses to a rf-only quadrupole lens.
Keywords: