A high-pressure liquid chromatographic-tandem mass spectrometric method for the determination of ethambutol in human plasma, bronchoalveolar lavage fluid, and alveolar cells
Authors:Conte John E  Lin Emil  Zhao Yeping  Zurlinden Elisabeth
Affiliation:Department of Epidemiology and Biostatistics, Infectious Diseases Research Laboratory, University of California, San Francisco, 94117, USA.
Abstract:A technique is presented for the specific and sensitive determination of ethambutol concentrations in plasma, bronchoalveolar lavage (BAL), and alveolar cells (AC) using a high-pressure liquid chromatographic (HPLC)-tandem mass spectrometric (MS-MS) method. The preparation of samples requires a deproteinization step with acetonitrile. The retention times for ethambutol, neostigmine bromide, and propranolol are 2.0, 1.4, and 1.1 min, respectively, with a total run time of 2.8 min. The detection limits for ethambutol are 0.05 microg/mL for plasma and 0.005 microg/mL for the BAL supernatants and AC suspensions. The assay has excellent performance characteristics and has been used to support a study of the intrapulmonary pharmacokinetics of ethambutol in human subjects.
Keywords: