Application of reference materials in the accurate calibration of the detection efficiency of a low-level gamma-ray spectrometry assembly for environmental samples
Authors:Y Nir-El
Institution:(1) Agency's Laboratories, International Atomic Energy Agency, A-2444 Seibersdorf, Austria;(2) Present address: Soreq Nuclear Research Centre, 81800 Yavn, Israel (permanent address)
Abstract:A new assembly dedicated for the low-level ψ-ray spectrometric measurements of environmental samples have been installed recently at the International Atomic Energy Agency's Laboratories, Seibersdorf. Calibration of the detection efficiency was performed by a set of 14 IAEA Certified Reference Materials, a standardized solution of134Cs, solutions of 2 pure potassium salts, and 2 mixed radionuclide gamma reference standard solutions. These materials, in 1.3 liter Marinelli beakers, were measured by a large HPGe detector. The high precision and accuracy of the calibration procedure is shown by the high significance achieved in the linear fits of the efficiency results, which are traceable to certified and standard reference materials. Small biases between some calibrands were detected. A new corrected result is the 1.37±0.05 Bq/kg activity concentration of137Cs in IAEA-A-14 Milk Powder.