首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Electrochemiluminescent determination of L-cysteine with a flow-injection analysis system.
Authors:Liande Zhu  Yingxiu Li  Guoyi Zhu
Institution:State Key Laboratory of Electroanalytical Chemistry, Changchun Institute of Applied Chemistry, Chinese Academy of Sciences, Changchun 130022, PR China.
Abstract:A flow-injection electrochemiluminescent method for L-cysteine determination has been developed based on its enhancement of the electrochemiluminecence of luminol at a glassy carbon electrode. This method is simple and sensitive for cysteine determination. Under the selected experimental parameters, the linear range for cysteine concentration was 1.0 x 10(-6) - 5.0 x 10(-5) mol/l, and the detection limit was 0.67 micromol/l (S/N = 3). The relative standard deviation for 11 measurements of 1.0 x 10(-5) mol/l cysteine was 4.5%. The proposed method has been applied to the detection of cysteine in pharmaceutical injections with satisfactory results.