首页 | 本学科首页   官方微博 | 高级检索  
     


On-line determination of manganese in solid seafood samples by flame atomic absorption spectrometry
Authors:M.C Yebra  A Moreno-Cid
Affiliation:Department of Analytical Chemistry, Nutrition and Bromatology, Faculty of Chemistry, University of Santiago de Compostela, Santiago de Compostela 15782, Spain
Abstract:Manganese is extracted on-line from solid seafood samples by a simple continuous ultrasound-assisted extraction system (CUES). This system is connected to an on-line manifold, which permits the flow-injection flame atomic absorption spectrometric determination of manganese. Optimisation of the continuous leaching procedure is performed by an experimental design. The proposed method allows the determination of manganese with a relative standard deviation of 0.9% for a sample containing 23.4 μg g−1 manganese (dry mass). The detection limit is 0.4 μg g−1 (dry mass) for 30 mg of sample and the sample throughput is ca. 60 samples per hour. Accurate results are obtained by measuring TORT-1 certified reference material. The procedure is finally applied to mussel, tuna, sardine and clams samples.
Keywords:Continuous ultrasound-assisted extraction system   Flame atomic absorption spectrometry   Manganese determination   Seafood
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏