On-line determination of manganese in solid seafood samples by flame atomic absorption spectrometry |
| |
Authors: | M.C Yebra A Moreno-Cid |
| |
Affiliation: | Department of Analytical Chemistry, Nutrition and Bromatology, Faculty of Chemistry, University of Santiago de Compostela, Santiago de Compostela 15782, Spain |
| |
Abstract: | Manganese is extracted on-line from solid seafood samples by a simple continuous ultrasound-assisted extraction system (CUES). This system is connected to an on-line manifold, which permits the flow-injection flame atomic absorption spectrometric determination of manganese. Optimisation of the continuous leaching procedure is performed by an experimental design. The proposed method allows the determination of manganese with a relative standard deviation of 0.9% for a sample containing 23.4 μg g−1 manganese (dry mass). The detection limit is 0.4 μg g−1 (dry mass) for 30 mg of sample and the sample throughput is ca. 60 samples per hour. Accurate results are obtained by measuring TORT-1 certified reference material. The procedure is finally applied to mussel, tuna, sardine and clams samples. |
| |
Keywords: | Continuous ultrasound-assisted extraction system Flame atomic absorption spectrometry Manganese determination Seafood |
本文献已被 ScienceDirect 等数据库收录! |
|