Sensitive and selective determination of molybdenum by flow injection chemiluminescence method combined with controlled potential electrolysis technique |
| |
Authors: | Jianxiu Du |
| |
Affiliation: | Department of Chemistry, School of Chemistry and Materials Science, Shaanxi Normal University, Xi’an 710062, PR China |
| |
Abstract: | A sensitive and selective flow injection chemiluminescence (CL) method combined with controlled potential electrolysis technique was described for the determination of molybdenum. The method is based on the chemiluminescence reaction of luminol with unstable molybdenum(III) in alkaline solution. The molybdenum(III) was on-line reduced from molybdenum(VI) via controlled potential electrolysis technique using a homemade flow-through carbon electrolytic cell at the potential of −0.6 V (versus Ag/AgCl). The method allows the determination of molybdenum in the 5.0×10−10 to 5.0×10−7 g ml−1 range with a limit of detection (3σ) of 5×10−11 g ml−1 molybdenum. The relative standard deviation is 2.6% for the 1.0×10−9 g ml−1 molybdenum solution in 11 repeated measurements. This method was successfully applied to the determination of molybdenum in water samples. |
| |
Keywords: | Chemiluminescence Flow injection Molybdenum |
本文献已被 ScienceDirect 等数据库收录! |