On-line preconcentration of palladium on alumina microcolumns and determination in urban waters by inductively coupled plasma mass spectrometry |
| |
Authors: | Mariella Moldovan M.Antonia Palacios |
| |
Affiliation: | Departamento de Qu?́mica Anal?́tica, Facultad de CC. Qu?́micas, Universidad Complutense de Madrid, 28040 Madrid, Spain |
| |
Abstract: | A method for on-line preconcentration of palladium at ultra-trace level on alumina microcolumns and determination by inductively coupled plasma mass spectrometry (ICP-MS) has been developed. A sampling time of 10 min (30 ml) and an eluent (KCN) volume of 300 μl provides a limit of detection of 1 ng l−1. The precision was 4% at the 25 ng l−1 level. The proposed system allows the on-line removal of the primary interferents (Cu and Y) providing the credible accuracy of the results. The proposed method is suitable for palladium determination in urban water samples. Platinum concentration in the analysed urban water samples was also determined. |
| |
Keywords: | Palladium Water Preconcentration Alumina Interferences ICP-MS |
本文献已被 ScienceDirect 等数据库收录! |