首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Analysis of zearalenone and α-zearalenol in animal feed using high-performance liquid chromatography
Authors:S De Saeger  L Sibanda
Institution:a Laboratory of Food Analysis, Faculty of Pharmaceutical Sciences, Ghent University, Harelbekestraat 72, 9000 Ghent, Belgium
b Toxi-Test n.v., Harelbekestraat 72, 9000 Ghent, Belgium
Abstract:Different extraction and clean-up techniques used before HPLC analysis were compared in order to obtain a reliable method for the quantitative determination of zearalenone (ZEA) and α-zearalenol (α-ZOL) in animal feed. Immunoaffinity clean-up was compared to C18 and Florisil column clean-up. Extracted samples were analysed by reversed-phase HPLC with fluorescence detection (λex=274 nm, λem=440 nm). A mobile phase of acetonitrile:water (50:50 (v/v)) and a flow-rate of 1.0 ml min−1 resulted in a good separation between ZEA and α-ZOL. Using immunoaffinity clean-up the linear range was between 25 and 600 μg kg−1 for ZEA and α-ZOL in maize. Intra-laboratory coefficients of variation (CV) (under repeatability conditions) were 9.16% for ZEA and 2.18% for α-ZOL. Recoveries for spiked ZEA and α-ZOL samples ranged from 89 to 110% with CVs between 5.2 and 11.2% (under within-laboratory reproducibility conditions). Using C18 and Florisil solid-phase clean-up, matrix interference was too high. Therefore, naturally contaminated animal feed samples were analysed using the developed HPLC method coupled to the immunoaffinity clean-up.
Keywords:Zearalenone  Animal feed  Clean-up  HPLC
本文献已被 ScienceDirect 等数据库收录!