首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Optimisation of a flow injection system with electrochemical detection using the desirability function: Application to the determination of hydroquinone in cosmetics
Authors:ME Rueda  A Herrero
Institution:a Department of Chemistry, Faculty of Sciences, University of Burgos, Plaza Misael Bañuelos s/n, Burgos 09001, Spain
b Department of Mathematics and Computation, Faculty of Sciences, University of Burgos, Plaza Misael Bañuelos s/n, Burgos 09001, Spain
Abstract:A procedure is proposed in which the determination of hydroquinone using a flow injection system with electrochemical detection is described. Size and coefficient of variation of the signal are optimised by a desirability function and a central composite design. The robustness of the optimum reached in the optimisation step is evaluated by means a Plackett-Burman design. The optimised FIA system is able to determine hydroquinone with a minimum detectable net concentration of 10 μg l−1 with a false positive probability of 0.05 and a false negative probability less than 0.05. In samples of bleaching cream, the proposed procedure has a recovery of 102.2% with standard deviation of 4.4% and a relative error of 6.2%.
Keywords:FIA  Electrochemical detection  Optimisation  Experimental design  Canonical analysis  Desirability functions  Robustness  Hydroquinone  Bleaching cream