首页 | 本学科首页   官方微博 | 高级检索  
     


On-line sorption preconcentration of metals based on mixed micelle cloud point extraction prior to their determination with micellar chemiluminescence: Application to the determination of chromium at ng l levels
Authors:Evangelos K Paleologos  Nicholaos P Evmiridis
Affiliation:Laboratory of Analytical Chemistry, Department of Chemistry, University of Ioannina, Ioannina 451 10, Greece
Abstract:The on-line incorporation of cloud point extraction (CPE) to flow injection analysis (FIA) is modified to extract and preconcentrate metal species rapidly, avoiding the formation of hydrophobic complexes, using a mixed micellar medium. Coupling the procedure with chemiluminescence (CL) detection based on the catalytic activity of metal species on the luminol-hydrogen peroxide reaction and enhancing the signal with the presence of a micellar carrier containing bromide ions produces a powerful tool for the preconcentration and determination of metal species at ng l−1 levels. As an analytical demonstration ultratrace concentrations of chromium were conveniently detected and quantified in samples with a complex matrix such as seawater and wastewater. The figures of merit for the determination of chromium were: 0.9-1.6% R.S.D. (n=5) with detection and quantification limits 0.5 and 2.0 ng l−1, respectively. The calibration graph was rectilinear from 2 to 200 ng l−1 (r=0.9996, n=6). Several other metal ions were determined in ideal situations, with analogous results.
Keywords:On-line   Cloud point extraction   Micellar-enhanced chemiluminescence detection   FIA manifold   Flow injection   Chromium(III)   Luminol
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号