Ketoconazol‐Triiodide Ion Pair Complex as a Suitable Carrier in an Iodide Selective Membrane Electrode
Authors:Ramin Maleki  Reza Hosseinzadeh  Amir Abbas Matin  Khalil Farhadi
Affiliation:1. Food & Chemical Analysis Research Lab., Jahad‐e‐Daneshgahi, Urmia, Iran;2. Departmant of Chemistry, Faculty of Science, Urmia University, Urmia, Iran
Abstract:A novel membrane sensor for selective monitoring of iodide, consisting of a triiodide‐ketoconazole ion pair complex dispersed in a PVC matrix, plasticized with a mixture of 2‐nitrophenyl octyl ether and dioctylphtalate with unique selectivity toward iodide ions, is described. The influence of membrane composition, pH of test solution and foreign ions on the electrode performance were investigated. The optimized membrane demonstrates a near‐Nernstian response for iodide ions over a wide linear range from 1.0 × 10?2 to 1.0 × 10?5 M, at 25 ± 1 °C. The electrode could be used over a wide pH range 3–10 and has the advantages of high selectivity, fast response time and good lifetime (over 4 months). It was successfully used as indicator electrode in potentiometric titrations and direct potentiometric assay of iodide ions.
Keywords:Ion‐selective electrode  Triiodide‐ketoconazole ion pair  Iodide  Potentiometry