首页 | 本学科首页   官方微博 | 高级检索  
     


Means of distinguishing selenium peaks from sulfur peaks in gas chromatography with a flame photometric detector
Authors:James R. Hancock  Christopher G. Flinn  Walter A. Aue
Affiliation:Department of Chemistry, Dalhousie University, Halifax Nova ScotiaCanada
Abstract:Flow optimization, optical discrimination, and doping with carbon disulfide or methane were tested as means of distinguishing selenium peaks from sulfur peaks in a flame photometric detector. Optical discrimination is the best choice if a dual-channel detector is available. For single-channel detectors, methane doping quenches the sulfur peak more severely than the selenium peak, thus allowing the peaks to be identified.
Keywords:
本文献已被 ScienceDirect 等数据库收录!