首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of residual isobutylene oxide--a genotoxic starting material in a drug substance by static headspace gas chromatography
Authors:Han Jun  Yeung David  Wang Fang  Semin David
Affiliation:Analytical Research & Development, Amgen Inc., Thousand Oaks, CA 91320, USA.
Abstract:A sensitive static headspace gas chromatography (sHS-GC) method is developed and validated for the determination of residual isobutylene oxide (IBO)-a genotoxic starting material in a drug substance for Phase I clinical trial studies. The experimental parameters, such as headspace vial pressure, headspace oven temperature, vial equilibration time, column flow rate, and GC oven temperature programs are optimized. Under the optimal conditions, the recovery is between 89.3% and 102.4% for spiked samples at three levels of IBO concentration in triplicates in sample preparations. Limits of quantitation (LOQ) and detection (LOD) of the standard solutions are 0.048 and 0.018 microg/mL, respectively. Linear range from 0.018 to 6 microg/mL is obtained with a correlation coefficient of 0.9999. The method is applied to determine residual IBO in drug substance samples from different batches.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号