Anodic stripping voltammetry with flow injection analysis |
| |
Authors: | Carolyn Wechter,Neal Sleszynski,John J. O Dea,Jane Osteryoung |
| |
Affiliation: | Department of Chemistry, State University of New York at Buffalo, Buffalo, NY 14214 U.S.A. |
| |
Abstract: | A flow cell with a wall-jet electrode design is used for anodic stripping voltammetry of lead at concentrations of about 10?7 mol dm?3. Maximum peak heights are obtained for narrow nozzle diameters and short nozzle-to-electrode distances. Linear calibration plots are obtainedfor almost four decades of change in concentration and can be extended by judicious choice of sample volume. Increasing sample throughput rates by increasing the solution flow rate decreases the analytical signal. Square wave voltammetry provides shorter analysis times and better sensitivity than differential pulse voltammetry. |
| |
Keywords: | |