首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Flow injection analysis for traces of zinc with immobilized carbonic anhydrase
Authors:Kunio Kashiwabara  Toshiyuki Hobo  Eigo Kobayashi  Shigetaka Suzuki
Institution:Department of Industrial Chemistry, Faculty of Technology, Tokyo Metropolitan University, 2-1-1 Fukasawa, Setagaya-ku, Tokyo 158 Japan
Abstract:A small column packed with immobilized bovine carbonic anhydrase is used for determination of traces of zinc in aqueous solution, based on the measurement of recovered esterase activity of the metal-free apoenzyme after taking up zinc from the sample solution. Conditions for the removal of zinc from the immobilized enzyme and for activity measurement are established. A linear calibration graph is obtained between 1 × 10?8 and 4 × 10?7 g of zinc. Then the method is successfully applied to the determination of zinc in tea, sediment and spring water.