Determination of arsenic at the picogram level by atomic absorption spectrometry with miniaturized suction-flow hydride generation
Authors:Masahiko Ikeda
Institution:Nippon Jarrell-Ash Company, 28, Joshunga mae-cho, Shimotoba, Fushimiku, Kyoto 612 Japan
Abstract:A continuous-flow hydride generator is modified and miniaturized for the determination of picogram amounts of arsenic by atomic absorption spectrometry. A 300-μl sample is dropped into a teflon cup and pumped into an alkaline sodium tetrahydroborate stream, which is acidified in a reaction coil. The evolved hydride is swept with argon through a phase separator into an electrically-heated quartz absorption cell and the absorbance is recorded. To eliminate differences in sensitivity between arsenic(III) and arsenic(V) without prereduction by potassium iodide, it is important that arsenic(V) be mixed with tetrahydroborate prior to mixing with hydrochloric acid. The method has a detection limit of 0.08 ng As ml?1 (24 pg) and the calibrations is linear up to 50 ng As ml?1. The relative standard deviation for 10 replicate measurements is 5.4% for 0.5 ng As ml?1. The addition of potassium iodide and hydroxylamine is confirmed to be effective in minimizing some interferences. The sampling rate is 90 h?1. Results for NBS biological and steel reference materials demonstrate applicability of the technique.