Enzymatic determination of glucose in a flow system by catalytic oxidation of the nicotinamide coenzyme at a modified electrode
Authors:Roger Appelqvist  György Marko-Varga  Lo Gorton  Arne Torstensson  Gillis Johansson
Institution:Department of Analytical Chemistry, University of Lund, P.O. Box 124, S-221 00 Lund Sweden
Abstract:A chemically modified electrode for detection of dihydronicotinamide adenine dinucleotide (NADH) and dihydronicotinamide adenine dinucleotide phosphate (NADPH) is described. Graphite rods were modified by dipping them into solutions of-dimethylamino-1,2-benzophenoxzinium salt (Meldola blue). The modified electrodes were mounted in a flow-through cell in a flow-injection manifold. Samples (50 μl) of pure nicotinamide coenzymes produced strictly linear calibration graphs from 1 μM to 10 mM with a repeatability of 0.2–0.6% RSD. A packed-bed enzyme reactor (210 μl) containing immobilized glucose dehydrogenase was inserted in the manifold for glucose determinations. Oxidized coenzyme was also added to the carrier electrolyte. Straight calibration graphs were again obtained up to 1mM β-d-glucose. The detection limit was 0.25 μM β-d-glucose for 50-μl samples. The electrode was kept at ?50 to 0 m V vs. SCE which was low enough to avoid interferences from ascorbic acid, uric acid or quinones.