The use of a cupric ion-selective electrode as end-point detector in precipitation titration of oxalate
Authors:Marco Taddia
Affiliation:(1) "ldquo"G. Ciamician"rdquo" Chemical Institute of the University, Via Selmi 2, I-40126 Bologna, Italy
Abstract:Summary It is demonstrated that oxalate can be titrated at pH 6–8 by employing lead(II) as titrant and a cupric ISE as end-point detector in the presence of traces of cupric ions. The response of the ISE to oxalate and E/pH plots have been examined. The method has been tested for the range of 0.8–40 mM oxalate. Recovery was found generally to be better than 98% and relative standard deviation was 0.4% at the 4 mM level. Interferences from carbonate, phosphate and arsenate can in part be overcome by adjusting the pH to 6.0±0.2.
Verwendung einer Kupfer-ionenselektiven Elektrode als Endpunkt-Detektor bei der Fällungstitration von Oxalat
Zusammenfassung Es wird gezeigt, daß Oxalat bei pH 6–8 mit Blei(II)-Lösung in Gegenwart von Kupferionen titriert werden kann, wobei eine Kupfer-ionenselektive Elektrode zur Endpunktsindikation dient. Das Verfahren wurde im Bereich 0,8–40 mM Oxalat getestet und die Anzeigeempfindlichkeit untersucht. Die Wiederfmdensrate ist besser als 98%; die relative Standardabweichung beträgt 0,4% bei 4 mM Oxalat. Durch Einstellung des pH-Wertes auf 6,0±0,2 können Störungen durch Carbonat, Phosphat und Arsenat z.T. eliminiert werden.
Keywords:Best. von Oxalat    llungstitration/Elektroden, ionenselektive  Cu-Elektrode zur Endpunktsindikation