A convenient method for routine estimation of dead time in gas chromatography
Authors:Watanachaiyong T  Jeyashoke N  Krisnangkura K
Affiliation:School of Bioresources and Technology, King Mongkut's University of Technology, Thonburi, Bangkok, Thailand.
Abstract:A set of retention times (tR) of n-alkanes at different temperatures and the primary dead times (tM) are used to determine the 4 numerical constants (a, b, c, and d) of an equation. This equation is rearranged into a second equation, used in turn for routine calculation of the secondary dead time (tMS) of each chromatogram from any member of the n-alkane series. Equation 2 can be used to calculate the tMS of both packed and capillary columns. The calculated tMS values are in good agreement with those of the tM values. The greatest difference is 3.38%, but it is speculated that the percent difference would be lower when more tR data are collected for the determination of the 4 numerical constants of Equation 1. Error in measurement of retention time undoubtedly would affect the accuracy of the estimated dead time, but it is attenuated by a factor of 1 + e(a + bN + c/T + dN/T).
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏