At-column injector for capillary GC — design and evaluation |
| |
Authors: | Gunnar Hagman Johan Roeraade |
| |
Abstract: | The present paper describes constructional details and evaluations of an at-column injector for capillary GC. Injections were made via a sample loop on a 0.32 mm i.d. capillary column. Two rotary valves were employed to allow a wash of the sample loop and a backflush of the transfer line. Repeatability, calculated from absolute area counts for n-alkanes was between 0.3–1% RSD, for injected sample volumes between 5 and 100 μl. Promising results were also obtained with syringe-based injections on narrow bore (100 μ i.d.) columns. Repeatability on the basis of normalized area counts was in the order of 0.1–0.2% RSD, while solvent tailing was practically absent. |
| |
Keywords: | Gas chromatography, GC Sample introduction At-column injector |
|
|