首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A photometry method for measuring tracks density on SSNTDs
Authors:Khalid M Abumurad  Adnan M Ismail  Hassan Abu-Safia
Institution:

Department of Physics, Faculty of Science, Yarmouk University, P.O. Box 556, 21163 Irbid, Jordan

Abstract:The purpose of this work is to determine alpha track density on solid state nuclear track detectors (SSNTDs) such as CR-39 efficiently with minimal time. Samples of CR-39 were irradiated with different doses from an alpha emitter. These samples were chemically etched and the tracks counted using an optical microscope. The intensity of the transmitted light (λ=400–1200 nm) through irradiated and etched CR-39 was measured using a photomultiplier. For all wavelengths used in this study, it is found that the transmittance of light decreases as irradiation time increases. In addition, the calibration factor was calculated from the relation between track density on CR-39 and light transmittance. The procedure appeared to be a good and fast technique for determining radon concentration in the environment and consequently dose reading.
Keywords:Photomultiplier tube  Transmittance  Absorbance  Counters  SSNTD