首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of fresh and old eluate of 99Mo/99mTc generators used for labeling of different pharmaceutical kits
Authors:N. Urbano  S. Modoni  M. Guerra  M. Chinol
Affiliation:(1) Division of Nuclear Medicine, European Institute of Oncology, Via G. Ripamonti, 435, 20141 Milan, Italy;(2) Department of Nuclear Medicine, Centro di Riferimento Oncologico della Basilicata, Rionero in Vulture (PZ), Italy;(3) Department of Nuclear Medicine, “Casa Sollievo della Sofferenza”, San Giovanni Rotondo (FG), Italy;(4) Division of Nuclear Medicine, European Institute of Oncology, Via G. Ripamonti, 435, 20141 Milan, Italy
Abstract:Summary Sixty 99Mo/99mTc wet column generators, loaded with two different 99Mo activities, were analyzed in order to assess the quality of their eluates. Each elution was used for labeling of different radiopharmaceuticals, in order to evaluate whether “risky” elutions, namely those performed just after generator delivery and at 72 hours or more from the last elution, could be conveniently employed when fresh available radioactivity is not enough for the planned labeling or when shipping problems arise, or delay in delivery of a new generator occurs. Radiochemical quality control of all radiopharmaceuticals labeled with these elutions was performed. The elutions differed mainly in 99Tc ground state (99gTc) and amounts of oxidizing impurities. Radiolabeling procedures, however, were not affected, suggesting that these “risky” elutions might be appropriately used, in “emergency” conditions, for labeling radiopharmaceuticals although their radiochemical purity control is recommended prior to patient administration.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号