Fitting formula for the injection volume of a gas chromatograph for radio‐xenon sampling in the lower troposphere |
| |
Authors: | Liu Shu‐jiang Chen Zhan‐ying Wang Shi‐lian Chang Yin‐zhong Li Qi Fan Yuan‐qing Zhao Yun‐gang Jia Huai‐mao Zhang Xin‐jun Wang Jun |
| |
Institution: | CTBT Beijing National Data Centre and Radionuclide Laboratory, Beijing, China |
| |
Abstract: | GC is usually used for xenon concentration and radon removal in the International Monitoring System of the Comprehensive Nuclear‐Test‐Ban Treaty. In a gas chromatograph, the injection volume is defined to calculate the column capacity. In this paper, the injection volume was investigated and a fitting formula for the injection volume was derived and discussed subsequently. As a consequence, the xenon injection volume exponentially decreased with the column temperature increased, but exponentially increased as the flow rate increased. |
| |
Keywords: | Atmospheric chemistry gas chromatography Injection volume Radio‐xenon sampling |
|