Qualification of automated low-field NMR relaxometry for quality control of polymers in a production setting
Authors:Sarah C. Chinn   Angela Cook-Tendulkar   Robert Maxwell   Hilary Wheeler   Mark Wilson  Z. Harry Xie
Affiliation:

aLawrence Livermore National Laboratory, 7000 East Avenue, Livermore, CA 94550, USA

bHoneywell Federal Manufacturing & Technologies, Kansas City Plant, Kansas City, MO 64141, USA

cBruker Optics, minispec Division, The Woodlands, TX 77381, USA

Abstract:Implementation of a low-field time-domain nuclear magnetic resonance (NMR) scanner as a diagnostic tool in the production of new polymer components is described in the context of qualification of a new QA/QC device. A study to determine the optimal experimental parameters was performed and a robotic autosampler was built to enable scanning of multiple pads. Relationships between T2 values and physical properties of DC745 slabs were investigated, and the appropriate sampling parameters for the production setting were determined. Two versions of a robotic autosampler were built and, for the component described here, a fourth radial axis was required in addition to traditional X, Y, and Z movement to eliminate the large variability in T2 due to inconsistent sample coverage caused by the complex rib geometry of the component. Data show that with appropriate choice of experimental conditions of the NMR detector and the detection geometry of the robotic autosampler, sufficient resolution of variations in cross-link density on the millimeter scale could be determined. All data to date demonstrate that low-field NMR devices are a feasible tool for use in production settings for non-destructive quality control of polymer components.
Keywords:NMR   Polymer testing   Quality control   PDMS   TD-NMR   Cross-link
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号