首页 | 本学科首页   官方微博 | 高级检索  
     


Mass measurement accuracy comparisons between a double-focusing magnetic sector and a time-of-flight mass analyzer
Authors:Bereman Michael S  Lyndon Matthew M  Dixon R Brent  Muddiman David C
Affiliation:W. M. Keck FT-ICR Mass Spectrometry Laboratory, Department of Chemistry, North Carolina State University, Raleigh, NC 27695, USA.
Abstract:We report a direct comparison of the mass measurement accuracies (MMAs) obtained on different mass spectrometry instrument types; a magnetic sector as the 'gold standard' and an electrospray ionization time-of-flight (ESI-TOF) instrument. Sixty samples, obtained from the Department of Chemistry at North Carolina State University, were analyzed on each instrument. Data are presented and compared between the different instruments. The average absolute MMAs achieved for the magnetic sector and Agilent ESI-TOF mass spectrometers were 3.0 and 1.1 ppm, respectively.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号