首页 | 本学科首页   官方微博 | 高级检索  
     


Liquid chromatographic determination of carotenoids and vitamins A and E in multivitamin tablets
Authors:Sun J
Affiliation:Enzymatic Therapy, Inc., Green Bay, WI 54311, USA.
Abstract:Carotenoids and vitamins A and E in multivitamin tablets can be determined simultaneously by reversed-phased liquid chromatography (LC) with a programmable UV detector. Samples were dissolved in dimethyl sulfoxide and then extracted with hexane. A portion was injected onto a Symmetry C18, 150 x 4.6 mm id, 5 microns column and chromatographed with a mobile phase of acetonitrile--0.25% ammonium acetate in methanol and 0.05% triethylamine in dichloromethane. A step gradient was used. The system was operated at 25 degrees C with a flow rate of 1.5 mL/min. UV detection was at 325 nm for retinols, 285 nm for tocopherols, and 450 nm for carotenoids. Detection limits were less than 0.3 ng for retinol and retinyl acetate; 2 ng for alpha-tocopherol acid succinate; 10 ng for alpha-tocopherol, gamma-tocopherol, and alpha-tocopherol acetate; and 0.4 ng for alpha-carotene and beta-carotene. Intraday and interday coefficients of variation ranged from 1.40 to 5.20%. The sample preparation method and LC assay are practical for quality control and routine analysis of multivitamin tablets.