首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Application of a hexapole collision and reaction cell in ICP-MS Part I: Instrumental aspects and operational optimization
Authors:I Feldmann  N Jakubowski  D Stuewer
Institution:Istitut für Spektrochemie und angewandte Spektroskopie (ISAS), Postfach 10 13 52, D-44013 Dortmund, Germany, DE
Abstract:The application of an ion-guiding buffer gas-filled hexapole collision and reaction cell in ICP-MS has been studied in order to give a preliminary performance characterization of a new instrument providing this feature for increasing the ion yield and decreasing contributions from Ar induced interfering molecular ions. As buffer gas He was used while H2 served as reaction gas. Addition of the latter can be an effective means for reduction of typical argon induced polyatomic ions (Ar+, ArO+, Ar2 +) by orders of magnitude owing to gas phase reactions. Molecular interferences generated in the cell can be suppressed by a retarding electric field established by a dc hexapole bias potential of –2 V.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号