首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Gas chromatographic retention indices of fentanyl and analogues
Authors:Manral Laxmi  Kumar Gupta Pradeep  Ganesan Kumaran  Chandra Malhotra Ramesh
Institution:Defence Research & Development Establishment, Gwalior-474002 (MP) India.
Abstract:Narcotic analgesics of the fentanyl class are characterized by high potency and relatively short duration of action. These compounds nowadays have become a substitute for heroine and are highly addictive for abusers. Herein, we report retention indices of fentanyl and its eighteen analogues relative to the homologous n-alkane series. These values are determined on a moderately polar BP-5 capillary column under programmed temperature and isothermal chromatographic conditions. The analogues differ in the substituent attached to the piperidine ring nitrogen, and retention indices are found to vary according to the nature of the substituent. The effects of chromatographic conditions like temperature programming rate, carrier gas flow rate, and oven temperature are studied. Retention indices are also determined on a non-polar BP-1 column to study the influence of stationary phase polarity. Standard deviation of all the RI values is less than one index unit.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号