首页 | 本学科首页   官方微博 | 高级检索  
     


A 3 × 6 arrayed CCD X‐ray detector for continuous rotation method in macromolecular crystallography
Authors:Kazuki Ito  Yurii Gaponov  Noriyoshi Sakabe  Yoshiyuki Amemiya
Abstract:A 3 × 6 arrayed charge‐coupled device (CCD) X‐ray detector has been developed for the continuous‐rotation method in macromolecular crystallography at the Photon Factory. The detector has an area of 235.9 mm × 235.9 mm and a readout time of 1.9 s. The detector is made of a 3 × 6 array of identical modules, each module consisting of a fiber‐optic taper (FOT), a CCD sensor and a readout circuit. The outputs from 18 CCDs are read out in parallel and are then digitized by 16‐bit analog‐to‐digital converters. The advantage of this detector over conventional FOT‐coupled CCD detectors is the unique CCD readout scheme (frame transfer) which enables successive X‐ray exposures to be recorded without interruption of the sample crystal rotation. A full data set of a lysozyme crystal was continuously collected within 360 s (180° rotation, 3 s/1.5° frame). The duty‐cycle ratio of the X‐ray exposure to the data collection time was almost 100%. The combination of this detector and synchrotron radiation is well suited to rapid and continuous data collection in macromolecular crystallography.
Keywords:rapid data collection  continuous‐rotation method  macromolecular crystallography  CCD X‐ray detector.