首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of acetaldehyde in human blood by high-performance liquid chromatography using fluorometry
Authors:M Takayanagi  S Goto  Y Kokubo  M Suzuki  T Yashiro
Abstract:A method was developed for the measurement of acetaldehyde in human blood by high-performance liquid chromatography (HPLC). The method was based on a pre-column reaction; a fluorescent substance was formed by a coupling reaction between 2 mol of cyclohexan-1,3-dione and 1 mol of acetaldehyde with ammonium acetate. The coupling compound was analyzed by HPLC and the concentration of acetaldehyde was obtained from the calibration curve drawn from the results using standard solutions. The determination was sensitive and reproducible with a range of 0.2-10 microM and a precision (coefficient of variation) of 2.43%.