首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of creatinine in human serum by short-end injection capillary zone electrophoresis
Authors:Zinellu Angelo  Carru Ciriaco  Usai Maria F  Sotgia Salvatore  Deiana Luca
Institution:University of Sassari, Sassari, Italy.
Abstract:A new ultra-rapid free-solution capillary zone electrophoresis method to measure serum creatinine is presented. Procedural parameters such as injection mode, concentration and pH of phosphate running buffer and acidic deproteinization of serum samples were investigated. Short-end injection permits a decrease of the analysis time by injecting samples at the outlet end of a silica capillary closest to the detection window, so reducing the migration distance. Thus, when a capillary with an effective length of 10.2 cm and a 40 mmol/L sodium phosphate buffer pH 2.35 was used, the obtained migration time of the creatinine peak was the shortest never described before, about 1.1 min. These conditions give a good reproducibility of the migration times (coefficient of variation, CV% < 0.5) and the peak areas (CV% < 2.8). Intra- and interassay CV were 3.06 and 6.26%, respectively, and analytical recovery was 99.4%. We compared our proposed method to Jaffé colorimetric assay, by measuring serum creatinine in 128 normal subjects. The obtained data were analyzed by the Passing and Bablok regression and Bland-Altman test. Creatinine concentration in healthy subjects was also used to investigate on its relationships with plasma thiols levels.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号