Automated gravimetric sample pretreatment using an industrial robot for the high-precision determination of plutonium by isotope dilution mass spectrometry. |
| |
Authors: | Naoki Surugaya Toshiaki Hiyama Masaru Watahiki |
| |
Institution: | Japan Atomic Energy Agency, 4-33 Muramatsu, Tokai, Naka, Ibaraki, Japan. surugaya.naoki@jaea.go.jp |
| |
Abstract: | A robotized sample-preparation method for the determination of Pu, which is recovered by extraction reprocessing of spent nuclear fuel, by isotope dilution mass spectrometry (IDMS) is described. The automated system uses a six-axis industrial robot, whose motility is very fast, accurate, and flexible, installed in a glove box. The automation of the weighing and dilution steps enables operator-unattended sample pretreatment for the high-precision analysis of Pu in aqueous solutions. Using the developed system, the Pu concentration in a HNO(3) medium was successfully determined using a set of subsequent mass spectrometric measurements. The relative uncertainty in determining the Pu concentration by IDMS using this system was estimated to be less than 0.1% (k = 2), which is equal to that expected of a talented analyst. The operation time required was the same as that for a skilled operator. |
| |
Keywords: | |
|
|