首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Detection of analyte binding to microarrays using gold nanoparticle labels and a desktop scanner
Authors:Han Anpan  Dufva Martin  Belleville Erik  Christensen Claus B V
Institution:BioArray Project, Mikroelektronik Centret (MIC), Technical University of Denmark (DTU), Oersteds Plads 345 East, DK-2800 Kongens Lyngby, Denmark.
Abstract:Microarray hybridization or antibody binding can be detected by many techniques, however, only a few are suitable for widespread use since many of these detection techniques rely on bulky and expensive instruments. Here, we describe the usefulness of a simple and inexpensive detection method based on gold nanoparticle labeled antibodies visualized by a commercial, office desktop flatbed scanner. Scanning electron microscopy studies showed that the signal from the flatbed scanner was proportional to the surface density of the bound antibody-gold conjugates, and that the flatbed scanner could detect six attomoles of antibody-gold conjugates. This detection system was used in a competitive immunoassay to measure the concentration of the pesticide metabolite 2,6-dichlorobenzamide (BAM) in water samples. The results showed that the gold labeled antibodies functioned comparably with a fluorescent based immunoassay for detecting BAM in water. A qualitative immunoassay based on gold-labeled antibodies could determine if a water sample contained BAM above and below 60-70 ng L(-1), which is below the maximum allowed BAM concentration for drinking water (100 ng L(-1)) according to European Union legislation.
Keywords:
本文献已被 PubMed 等数据库收录!