首页 | 本学科首页   官方微博 | 高级检索  
     


Application of hydride generation to the determination of trace concentrations of arsenic by inductively-coupled plasma atomic emission spectrometry
Authors:Taketoshi Nakahara
Affiliation:Department of Applied Chemistry, College of Engineering, University of Osaka Prefecture, Sakai, Osaka 591 Japan
Abstract:A method is described for the determination of traces of arsenic, based on sodium tetrahydroborate reduction of arsenic and introduction of the arsine formed to a relatively low-powered (1.6 kW) inductively-coupled plasma where the arsenic atomic emission is generated, Inter-element interference effects are described; many can be eliminated by addition of potassium iodide to the sample solutions, but the standard additions method is recommended for accurate arsenic determination. Potassium iodide is also used for prereducing arsenic(V) to arsenic(III). The method has a detection limit of 0.1 ng As ml-1 and the linear calibration range spans nearly four orders of magnitude. The proposed method is applied to the determination of ng ml-1 levels of arsenic in waste waters.
Keywords:
本文献已被 ScienceDirect 等数据库收录!